P/ACE™ MDQ Series Capillary Electrophoresis Systems Technical specifications
Modes of Operation Constant/Gradient Voltage†
Constant/Gradient Current†
Constant/Gradient Power†
Variable pressure
Sample Introduction Pressure, Vacuum, Electrokinetic
Voltage Range 1 - 30 kV (1 - 10 kV for injections),
Programmable in 0.1 kV increments
Reversible voltage programmed through user interface.
Current Range 3 - 300µ Amps
Pressure Range -5 to +100 psig (Rinse/Separations); -5 to +25 psig (Sample Introduction)
Sample Temperature 5° C to 60° C
Sample temperature environment is independent of the ambient buffer environment.
System Capacity
Sample Tray 2 x 96-well plates
2 x 48 2 mL vials, 2 x 48 0.5 mL vials, 2 x 48 PCR vials
Buffer Tray 2 x 36 2 mL vials, 4 x 25 mL trays
Capillary Cartridge Recirculating liquid coolant 15° to 60° C
Wavelength Range Diode Array 190 - 600 nm
UV/Vis 200, 214, 254, 280 nm standard filter
190 - 600 nm (with custom filter options)
Detectors Diode Array
UV/Vis
Laser-Induced Fluorescence
Scan Collection Diode Array 0.5 - 32 Hz (user selected)
Data Rate Collection 0.5 to 32 Hz
Electrical Voltage 90 - 240 V 50/60 Hz